Print

A specialist gas analysis system for fast event transient analysis

Measures the concentration of gases and vapours from low PPM to high % levels in < 0.2 seconds.

The Hiden HPR-20 QIC TMS Transient MS is configured for fast event gas analysis of gases and vapours at pressures near atmosphere. Ideal for fast gas switching experiments the MS features the Hiden QIC quartz-lined 0.9 m sampling interface.

Overview

The inlet, operating at 200 °C, provides response times of less than 150 ms to changes in gas composition with a 5 decade response time in < 200 ms.

The QIC inlet is coupled directly to Hiden’s Pulse Ion Counting (PIC) digital MS which is capable of measurement speeds of up to 500 data points/s over the entire 7 decade dynamic range.

Features

- 0.9 m fast response QIC capillary - 150 ms response time

- Open ion source and optimised pumping configuration for fast response

- Digital PIC detector - 7 decades continuous log scale

- Detection of low ppm to high % levels in < 200 ms

Specifications

Mass range

200 or 300 amu

Minimum detectable concentration

5 PPB

Maximum detectable concentration

100%

Sample inlet

Heated flexible silica capillary

Standard capillary length

0.9 meters

Sample inlet pressure

100 mbar to 2 bar abs

Digital pulse ion counting detector 

7 decade continuous log scale

Response time to gas concentration change

150 milliseconds

Response time to 5 decade change

< 200 ms

Fast measurement speed

Up to 500 measurements per second 

Sample flow required (Standard)

16 atm cc/min

Sample flow required (Optional)

Options for < 1 atm cc/min to > 20 atm cc/min

Corrosive gas sampling options

Yes